مشخصات فني


مولتی متر دیجیتالی هیوکی 3801-50 , 3802-50
قابلیت اندازه گیری ولتاژو جریان ac/dc و درجه حرارت وقابلیت ثبت مقادیر ماکزیمم و مینیمم و میانگین توسط حافظه داخلی
 

DIGITAL HiTESTER 3801-50, 3802-50

CE
CAT III 1000 V
CAT IV 600 V
 
 
High Performance, Superior Functionality, and Advanced Safety
 
 High-performance model 3801-50 (AC+DC mode/ Temperature/ Dutyratio measurement)
Can record to internal memory, maximum, minimum, or average value
 High-precision measurement at the 3801-50 Basic accuracy ±0.025 %, 1 μV minimum resolution
Depend on a dual-display, can display two parameter
Convenient back-light display on dark environment 
 
 
مولتی متر دیجیتالی هیوکی 3801 / 3802
 3801-503802-50
DC Voltage /51000 dgt.51.000 mV to 1000.0 V, 7 ranges51.000 mV to 1000.0 V, 7 ranges
Basic accuracy: ±0.025 % rdg. ±5 dgt.Basic accuracy: ±0.03 % rdg. ±5 dgt.
AC Voltage /45 to 1 kHz51.000 mV to 1000.0 V, 7 ranges
Basic accuracy: ±0.4 % rdg. ±25 dgt.Basic accuracy: ±0.6 % rdg. ±25 dgt.
Frequency characteristics: 20 Hz - 100 kHzFrequency characteristics: 30 Hz - 30 kHz
Resistance510.00 Ω to 510.00 MΩ, 510.00 nS, 8 ranges (3802-50: up to 51.000 MΩ, 510.00 nS, 7 ranges)
Basic accuracy: ±0.05 % rdg. ±5 dgt.Basic accuracy: ±0.08 % rdg. ±5 dgt.
DC Current510.00 μA to 10.000 A, 6 ranges
Basic accuracy: ±0.05 % rdg. ±25 dgt.Basic accuracy: ±0.1 % rdg. ±25 dgt.
AC Current /45 to 2 kHz510.00 μA to 10.000 A, 6 ranges
Basic accuracy: ±0.7 % rdg. ±20 dgt.Basic accuracy: ±0.8 %rdg. ±20 dgt.
Frequency characteristics: 20 Hz - 100 kHzFrequency characteristics: 30 Hz - 20 kHz
C (capacitance)9.999 nF (resolution 1 pF) to 99.99 mF (rosolution 10 μF), 8 ranges
Basic accuracy: ±1.5 % rdg. ±5 dgt.
Continuity buzzer1000 digits or less at resistance measurement
Diode checkOpen terminal voltage: 4.8 V, Testing current: 1 mA
Hz (Frequency, Volt function)99.999 Hz to 999.99 kHz 5 ranges, ±0.02 % rdg. ±3 dgt.
Hz counterDivide 1: 99.999 Hz to 9.9999 MHz
Divide 100: 9.9999 MHz to 99.999 MHz
Not available
Accuracy: ±0.002 % rdg. ±3 dgt.
Duty ratio99.99 % , Basic accuracy: ±0.3 %/kHz ±0.3 %
Pulse width510.00 ms to 1999.9 ms, Accuracy: ±0.2 % rdg. ±3 dgt. at 10 μs over pulse width
Temperature measure-200.0 °C to 1372.0 °C (Thermo-couple K), -210.0 °C to 1200.0 °C (Thermocouple J), Accuracy: ±0.3 % rdg. ±3 °C add probe accuracy, (the 3802-50: K type only)
Pulse output0.5 Hz to 4800 Hz, change duty ratio possibleNot available
Other functionAC+DC modeNot available
dBm mode, Recoding (Max./ Min. / Ave.), Relative value, Peak hold, Trriger hold, Auto-power save, Terminal A improper connection warning, Back light
DisplayDigital/ LCD, max. 51000 digits
Sample rate3.75 times/s (V and, A measurement)
Power supplyStacked battery (6F22 or 6LR61) ×1, Continuous use: 50 hours (6LR61)
Dimensions and mass100W (3.94in) × 202H (7.95in) × 57D (2.24in) mm, 680 g (24.0 oz) (with holster and battery)
Supplied accessoriesTEST LEAD×1, Strap×1, Holster×1, Instruction Manual×1, Stacked Alkaline Battery (6LR61)×1 (for trial purposes only)
 کاتالوگ.jpg